Radiation protection for caregivers in projection X-ray imaging

Loading...
Thumbnail Image

Date

2023-12

Authors

Μαρκάτου, Μαίρη

Journal Title

Journal ISSN

Volume Title

Publisher

Abstract

In pediatric radiography, immobilization of infants is crucial for the avoidance of unnecessary patient doses. Physical holding by caregivers has proven to be highly effective and is commonly used. This study aimed to quantify the caregiver's exposure to scatter radiation while immobilizing a pediatric patient during a chest X-ray. The main objective was to measure the scatter radiation dose at different heights with respect to the floor or patient table at two usual locations where a caregiver may stand. An AGFA DX-D 600 X-ray was used, and a PMMA phantom was irradiated on the examination table. An RTI survey meter was fixed at four different points corresponding to the gonadal and thyroid gland areas of an adult caregiver: laterally from the phantom, 0.5 m away from the central beam, at 0.8 and 1.5 m heights with respect to the floor, and cranially, 0.4 m away from the central beam, at 0.05 m and 0.6 m heights from the patient table. The phantom was exposed at 65 kVp, 75 kVp, 85 kVp, and 95 kVp at a constant tube current-exposure time product of 2 mAs with a SID of 110 cm. The highest scatter radiation dose, 0.58 μGy, was recorded in the cranial position at 0.6 m height from the table for 95 kVp. This value is far below the dose constraints for caregivers, as recommended by the ICRP. However, the LNT model suggests that even the smallest dose could cause cancer, so measures should be taken to maintain the scatter radiation dose as low as reasonably achievable (ALARA), such as lead aprons or selection of the lateral position when possible.

Description

Keywords

Pediatric radiology, Chest X-ray, Caregivers, Scatter radiation dose, Radiation protection

Citation